The MAVOLUX COMPACT precision luxmeter has evolved through collaboration with leading cinematographers and lighting technicians of Germany’s public television broadcasters. One-hand operation provides leeway for simultaneous operation of communication devices and remote controls. The swiveling sensor permits adjustment to individual working conditions and the display can be read in the dark thanks to illumination in the HOLD function. With its carrying strap, the meter hangs securely around the operator’s neck and the included neoprene case offers protection during transport.
The MAVOLUX COMPACT can of course also be used in industrial settings for planning, installing, inspecting and monitoring lighting systems, as well as for the assurance of compliance with specified lighting conditions.
Classified measurement of illuminance in lx or fc in accordance with class C per DIN 5032-7, appendix B of IEC 13032-1 and CIE 69.
Accuracy amounts to + 3% + 1 digit of the display value.
100 measured values can be saved and retrieved.
High initial sensitivity and a resolution of 0.1 lx or 0.01 fc, right on up to large illuminance values of 199,900 lx or 19,990 fc.
As an option, the accredited GOSSEN Light Lab can issue a factory or a DAkkS calibration certificate for measuring equipment monitoring in accordance with DIN EN ISO 9001.
Simple one-hand operation, easy to read display with backlight in HOLD mode, swivel-head, compact design.
The spectral sensitivity of the silicon photodiode is color corrected and corresponds to the spectral brightness sensitivity of the human eye V(λ).
The luminosity of a flat measuring surface is proportional to the cosine of the incident angle of light. This relationship is taken into consideration by the receiver during evaluation.
Class C DIN 5032-7 / EN 13032-1 annex B
MAVOLUX COMPACT | |
---|---|
Illuminance [Lux] | 0.1 lx … 199 900 lx / 0.01 fc … 19990 fc |
Measuring ranges | 4 |
Measuring range selection | Automatic / Manual |
Measuring rate | 2 per second |
Measuring sensor | Silicon photodiode with V (λ) filter |
Measured value memory | 100 measured values |
Error limit – V( λ) adapted (f1′) typical | < 7.5 % |
Error limit – overal error typical | ≤ 15 % |
Accuracy | ± 3 % of reading + 1 digit |
MAVOLUX COMPACT | |
---|---|
Battery | 2 x 1.5V mignon , type AA |
Automatic battery control | • |
Automatic shutdown | 4 min / continuous operation |
Battery service life | Approx. 45 h with alkaline manganese battery |
MAVOLUX COMPACT | |
---|---|
Display | 3 1/2 digit LCD |
Back-Light | • |
Operating elements | 6 keys |
MAVOLUX COMPACT | ||
---|---|---|
Operating temperature | 0 °C to 50 °C | |
Dimensions | 65 mm x 140 mm x 25 mm | |
Weight | 105 g without battery | |
Delivery contents | Case, carrying strap, battery, operating instructions, calibration protocol | |
• Included function |
GOSSEN Foto- und Lichtmesstechnik GmbH . Lina-Ammon-Str. 22 . 90471 Nürnberg
T +49 911 800621 0 . F +49 911 800621 29 . info@gossen-photo.de